Dosage selector
This test is made to check that there are no internal leakage between dosage volumes and dosage selector.
- Make sure the apparatus is depressurised and the pressure is relieved (reserve gas valve open, by-pass performed).
- Set the dosage selector in the RED position.
- The dosage pressure gauge should still be mounted.
- Pressurise the IS-MIX.
- Check the dosage pressure gauge after 1 minute. Approved value: 0+0.1 bar.
- Remove the dosage pressure gauge and refit the plug on the dosage volume.